119
Load Correction (Correcting Values to Match Reference Values)
To reset the correction condition settings
This section describes how to clear all settings for the selected correction condition number.
Screen display method (For more information, see p. 27):
(Measurement screen)
ADJ
key>(
ADJ
screen)
LOAD
key
1
Select the correction condition number
to reset with the
keys and then
touch the
EDIT
key.
2
1
2
Touch the
RESET
key and then touch
the
SET
key.
Touch when you wish to cancel the reset.
(This dialog box will be closed.)
1
2
3
Touch the
EXIT
key twice.
Displays the measurement screen.
When load correction fails to complete normally
If correction fails, a window like the following appears. Touch
EXIT
to close the window and then
configure the correction conditions again.
5
Error Correction