184
External I/O Settings
Outputting measured values (switching to BCD mode) (LCR mode only)
This section describes how to switch the output mode (between judgment mode and BCD mode)
during LCR mode operation. The default setting is judgment mode, in which case
judgment results
are output. Changing the mode to BCD mode causes
measured values to be output.
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
IO BCD
key
1
Touch the
ON
key.
OFF
Sets the output mode to judgment
mode. (Judgment results will be
output.)
ON
Sets the output mode to BCD
mode. (Measured values will be
output.)
2
Set the decimal point position with the
keys.
PARA1
Sets the decimal point position for
the No. 1 parameter.
PARA3
Sets the decimal point position for
the No. 3 parameter.
(Select: 9.99999G/
999.999M/ 99.9999M/ 9.99999M/
999.999k/ 99.9999k/ 9.99999k/
999.999/ 99.9999/ 9.99999/
999.999m/ 99.9999m/ 9.99999m/
999.999µ/ 99.9999µ/ 9.99999µ/
999.999n/ 99.9999n/ 9.99999n/
999.999p/ 99.9999p)
Example:
To display the No. 1 parameter measured
value as 12.3456 µF and the No. 3 param
-
eter measured value as -80.567°, use the
following settings:
PARA1
:
99.9999
µ
PARA3
:
99.9999
3
Touch the
EXIT
key twice.
Displays the measurement screen.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...