121
Correcting Measured Values with a User-specified Correction Coefficient (Correlation Correction)
5.7
Correcting Measured Values with a User-specified
Correction Coefficient (Correlation Correction)
This functionality enables you to correct measured values using a user-specified correction coef
-
ficient. This function can be used to provide compatibility among measurement devices.
Set the correction coefficients A and B for the measurement values of No. 1 to
No. 4 parameters to correct by the expression on the right.
(See "Appx. 1 Measurement Parameters and Calculation Formula" (p. Appx.1).)
Y = A × X + B
However, if the parameter corresponding to X is either D or Q, scaling is applied
to
θ
as shown in the expression on the right, and then D or Q is obtained from
θ
'.
θ
' = A ×
θ
+ B
X: No. 1 or No. 3 parameter measurement value Y: the last measurement value
θ
': correction value of
θ
A: integration value of the measured value X
B: the value added to measured value X
Screen display method (For more information, see p. 27):
(Measurement screen)
ADJ
key>(
ADJ
screen)
SCALE
key
1
Touch the
ON
key, and then touch the
EXIT
key.
1
2
When you wish to cancel scaling
Touch
OFF
.
2
Select the correction coefficient of the
parameter you want to change.
The parameters and correction coefficient
numbers correspond as shown below.
SCALE1
No. 1 parameter
SCALE2
No. 2 parameter
SCALE3
No. 3 parameter
SCALE4
No. 4 parameter
3
Touch the
A
key.
Touch when you wish
to revert the setting to
the default value.
Touch when you wish
to cancel the setting.
4
Set correction coefficient A with the
numeric keypad and then touch the
EN-
TER
key.
1
2
Settable range: -999.999 to 999.999
If you make a mistake, touch the
C
key to
reenter the value.
Touching
ENTER
while nothing is being dis-
played (while touching the
C
key) will close
the dialog box without changing the setting.
The display will return to the screen shown
in step 3.
5
Error Correction
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...