210
Measurement Range and Accuracy
(2) Calculate the basic accuracy for capacitance Cs=160 nF.
(If the measurement conditions are a measurement frequency of 1 kHz and a measurement speed
of SLOW2)
Excerpted from the "Basic accuracy" (p. 207).
Z
θ
Range
1.0000 kHz to
10.000 kHz
100 kΩ
10 kΩ
A= 0.05
A= 0.03
B= 0.02
B= 0.02
1 kΩ
1
Measure the sample’s
Z
and
θ
values using auto-ranging.
Assume that the measured Z and
θ
values are as follows:
Z
=1.0144 k
Ω
,
θ
=-78.69°
Because
Z
is 1.0144 kΩ, 10 kΩ measurement range will be used.
2
Find the cell at the intersection of the row for the 10 kΩ range and the 1.0000 kHz to
10.000 kHz column (since the measurement frequency is 1 kHz) in the "Basic accuracy"
3
Calculate the basic accuracy for
Z
using the
Z
coefficients A and B.
Based on the accuracy table, coefficient A = 0.05 and coefficient B = 0.02
Plug in the values for the 1 k
Ω
range in the "Basic accuracy" (p. 207) into the follow-
ing equation:
Z
accuracy =
10 10.144
0.05 0.02
1
10
k
k
×
Ω
±
+
×
−
Ω
0.05% rdg.
4
Similarly, calculate the basic accuracy of
θ
.
The accuracy table yields the values A=0.03 and B=0.02
Using the basic accuracy formula of "Basic accuracy" (p. 207) for "1 k± or more".
θ
accuracy =
10 10.144
0.03 0.02
1
10
k
k
×
Ω
±
+
×
−
Ω
±0.03°
5
Calculate the range within which
Z
and
θ
values can be acquired from the basic accuracy.
0.05
min 1.0144
1
1.0139
100
0.05
max 1.0144
1
1.0149
100
min
78.69 0.03
78.72
max
78.69 0.03
78.66
Z
k
k
Z
k
k
θ
θ
=
Ω× −
Ω
=
Ω× +
Ω
= −
−
= −
°
= −
+
= −
°
≒
≒
6
Calculate the range within which Cs values can be acquired based on the
Z
and
θ
ranges.
(For more information about the Cs
calculation formula, see "Appx. 1 Measurement Param-
eters and Calculation Formula" (p. Appx.1).)
3
3
3
3
10×1.0144×10
Z
=
0.05+0.02×
-1
0.05%
10×10
10×1.0144×10
=
0.03+0.02×
-1
0.03
10×10
0.05
min 1.0144 × 1-
1.0139
100
0.05
max 1.0144 × 1-
1.0149
100
min
78.69 0.
Z
k
k
Z
k
k
θ
θ
±
±
±
±
°
=
Ω
Ω
=
Ω
Ω
= −
−
基本確度
≒
基本確度
≒
≒
≒
03
78.72
max
78.69 0.03
78.66
1
min
159.90nF
... 0.0625% rdg.
×Zmax×sin min
1
max
160.10nF
... 0.0625% rdg.
×Zmin×sin max
2
Cs
Cs
f
θ
ϖ
θ
ϖ
θ
ϖ
π
= −
°
= −
+
= −
°
=
−
=
= × ×
≒
≒
f
is frequency [Hz].
Consequently, the Cs basic accuracy is ± 0.0625% rdg.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...