101
Open Correction
Correction range limitation function (to shorten the correction time)
In All correction, correction is performed for the entire frequency range. By setting the minimum and
maximum correction frequencies with this function, you can reduce the time required to perform the
correction process. The DC on/off setting as well as the correction minimum and maximum fre
-
quency settings apply to both open and short correction.
For more information about the sequence of screens until the
AREA
key is displayed, see "All cor
rection" (p. 100), and (p. 106).
1
Touch the
AREA
key.
2
Select DC correction.
Touch when you wish
to cancel the configu
-
ration process.
Touch when you wish
to revert the settings
to their default values.
ON
Performs correction for AC measure
-
ment and DC measurement.
OFF
Performs correction for AC measure
-
ment only.
3
Touch the
MIN
key and enter the correc-
tion minimum frequency with the nu-
meric keypad.
1
2
Settable range: 4 Hz to 8 MHz
(default setting: 4 Hz)
If you make a mistake, touch the
C
key to
reenter the value.
Touch the
MIN
key to return to the previous
screen without changing the settings.
4
Press a unit key to confirm the setting.
•
The frequency is not confirmed until a unit
key is pressed.
•
If you attempt to set a measurement fre
-
quency greater than 8 MHz, it will automati
-
cally be reduced to 8 MHz.
•
If you attempt to set a measurement fre
-
quency lower than 4 Hz, it will automatically
be increased to 4 Hz.
The display will return to the screen shown
in step 2.
5
Touch the
MAX
key and enter the cor-
rection maximum frequency with the
numeric keypad.
Settable range: 4 Hz to 8 MHz
(default setting: 8 MHz
)
x
Correction cannot be performed if limits are
applied using a range that exceeds the maxi-
mum valid frequency setting (see p. 215) for
each cable length. Correction will be per-
formed up to the maximum valid frequency
setting for the set cable length if limits are
applied using a range that exceeds the maxi-
mum valid frequency setting (see p. 215) for
each cable length.
5
Error Correction
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...