External Input/Output Connector and Signals
Input (IN) signal function details
This section describes input (IN) signals.
Input (IN) sig-
nal
Detailed description
TRIG
―
―
―
―
•When the trigger setting is the external trigger (
EXT
), measurement is performed once
with the falling (
DOWN
) or rising (
UP
) edge of the TRIG
―
―
―
―
. The edge direction can be set in
the
SET
screen. (Initial value: Falling [
DOWN
])
See "Disabling the trigger input during measurement and setting the trigger input effective
edge" (p. 182). (Falling: HI→LO, rising: LO→HI)
•
The TRIG signal is invalid when the trigger source is set to the internal trigger (
INT
).
See "Trigger (perform measurements with user-defined timing) (Common)" (p. 63).
•You can set whether to enable or disable
TRIG
―
―
―
―
input during measurement (during output
of the
EOM
―――
(HI)).
"Disabling the trigger input during measurement and setting the trigger input effective
LD0
―
―
―
to LD6
―
―
―
Selects the number of the panel to load.
If a trigger signal is input in external trigger mode, the selected panel is loaded and used
for measurement.
Input the panel value as a binary value to
LD0
―
―
―
to LD6
―
―
―
.
<Example> OFF: HI (5 V to 24 V), ON: LO (0 V to 0.9 V)
Pin No.
LD6
――
―
LD5
――
―
LD4
――
―
LD3
――
―
LD2
――
―
LD1
――
―
LD0
――
―
Panel 1
OFF
OFF
OFF
OFF
OFF
OFF
ON
Panel 2
OFF
OFF
OFF
OFF
OFF
ON
OFF
Panel 3
OFF
OFF
OFF
OFF
OFF
ON
ON
Panel 4
OFF
OFF
OFF
OFF
ON
OFF
OFF
Panel 5
OFF
OFF
OFF
OFF
ON
OFF
ON
Panel 6
OFF
OFF
OFF
OFF
ON
ON
Panel 7
OFF
OFF
OFF
OFF
ON
ON
ON
Panel 8
OFF
OFF
OFF
ON
OFF
OFF
OFF
…
Panel 32
OFF
ON
OFF
OFF
OFF
OFF
OFF
…
Panel 127
ON
ON
ON
ON
ON
ON
ON
Panel 128
OFF
OFF
OFF
OFF
OFF
OFF
OFF
C1
―
―
,
C2
―
―
•C1
―
―
: Switches between the high-order digit and the low-order digit (exponent or decimal
point) in BCD mode.
OFF: High-order digit output; ON: low-order digit output (polarity, ERR)
•C2
―
―
: Switches between the No. 1 parameter and the No. 3 parameter in BCD mode.
OFF: No.1 parameter; ON: No. 3 parameter
•
For more information about BCD mode, see "BCD mode function details" (p. 168).
LD_VALID
―
― ―
―
―
―
―
•
Input a negative logic signal from an external source when performing a panel load op
-
eration.
•
Maintain LO level after TRIG
―
―
―
―
input until
INDEX
―
―――
―
is output.
CALIB
――
―
―
―
•
When the DC adjustment function is set to OFF during DC resistance measurement,
acquires the offset value generated by the internal circuitry at the user-defined timing.
•
Maintain LO level after TRIG
―
―
―
―
input until
INDEX
―
―――
―
is output.
9
External Control
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...