183
External I/O Settings
Setting the E
________
OM output method and output time
The higher the measurement frequency, the shorter the time that INDEX
―
―――
―
and EOM
―――
are HI (during
measurement).
If the HI time when receiving the INDEX
―
―――
―
and EOM
―――
signals is too short for reasons related to the
input circuit, you can configure the instrument so that the LO signal is held (i.e., so that
EOM
―――
output
is continued) after
EOM
―――
changes to LO when measurement completes before reverting to HI.
The
INDEX
―
―――
―
output method can be changed in the same manner.
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
IO EOM
key
1
Select the EOM
―――
output method.
HOLD
EOM
―――
will be in the HI state for
(analog measurement time +
calculation time + delay time t1)
Timing (Timing Charts)" (p. 171)).
PULSE
EOM
―――
will be in the HI state except
during the set pulse width time.
1
2
2
Set the EOM
―――
output time during
PULSE
operation with the
keys.
Settable range: 0.0001 s to 0.9999 s
If you make a mistake, touch the
C
key to
reenter the value.
•
For a timing chart illustrating operation
when using the
HOLD
and
PULSE
settings,
see p. 173.
•The output time can only be set when using
the
PULSE
output method. (The output time
when using the
HOLD
output method is the
time until output turns off after the receipt of
the next trigger, a minimum of 400 ms.)
3
Touch the
EXIT
key twice.
Displays the measurement screen.
9
External Control
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...