Note
Pay special attention to lens, edges, and groves.
4.
Turn the probe and continue wiping until the entire surface of the probe has been
cleaned. As the wipe becomes visibly soiled, discard the wipe into clinical trash and
dispense fresh wipes as needed.
5.
Wrap a clean wipe around a soft nylon bristle brush to access crevasses, such as biopsy
notches, on the surface of the probe.
6.
Visually inspect the probe for any remaining soil and, if necessary, repeat steps 3
through 5 until the probe is visibly clean.
Probe Manual Cleaning Instructions – Enzymatic Detergent
1.
Ensure the probe has been disconnected from the console. Put on a clean pair of gloves
and fill a sink or basin with warm potable water (30 - 40°C) to a level allowing immersion
of the probe up to the immersion line shown in the Instructions for Use .
2.
Prepare the cleaning solution in accordance with the detergent manufacturer’s
instructions.
3.
Immerse the probe in the prepared cleaning solution up to the immersion line and ensure
no air bubbles are trapped on the surface.
Caution
DO NOT submerge probe beyond the immersion line shown in the Instructions for Use .
4.
Brushing with a clean, soft, nylon bristle brush from the base of the cable strain relief to
the distal tip is critical to ensure cleaning and disinfection efficacy.
Figure 5-4 Cleaning the probe using a brush
Caution
Do not use the brush on the probe lens.
Figure 5-5 Probe Lens Examples
5.
Continue brushing the probe for not less than the minimum contact time listed on the
detergent manufacturer’s label.
Probes and Biopsies
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
5-9