Acceptance criteria for each probe
In the following table values for all probe types are listed that would cause a failed transducer
element check test. Defects less than the maximum number of defect elements or nearby
elements and larget gaps are rated asw passed.
Probes and Biopsies
5-32
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4