Disconnecting a probe
1.
Freeze the image by pressing Freeze.
Do not disconnect an active probe. Ensure that the probe is in
Freeze mode
before
disconnecting the probe, or open the Probe Selection menu and select a different probe.
Damages may occur.
2.
Push the probe lock-handle knob to right direction and remove the probe.
Entering patient data
1.
Select "Tap here to begin exam" or the menu bhtton on the top left corner of the main
screen.
2.
Press
Patient
on the left conlumn of the menu screen.
3.
Insert first and last name of the patient.
4.
Select the desired exam category.
5.
Fill in the gaps.
Changing the user presets
1.
Slide up probe menu in the bottom left of the touch monitor.
2.
Select the desired preset on the bottom left side of screen.
3D/4D scanning
1.
Select a 3D/4D volume probe.
2.
Optimize specific structures by using the available presets.
3.
Adjust the size and position of the ROI Box (Region of Interest) by pressing Change (top
trackball key). Press
Change
again or using on screen gestures.
4.
Adjust the volume acquisition by sliding the Volume Angle control on the left user
interface. The Volume Angle is displayed on the lower section of the monitor.
5.
Start the volume acquisition by pressing Start (right trackball key) or Freeze.
Info
For more information see Chapter 8.
Using the Freeze button
1.
Press Freeze to freeze the image.
2.
Pay attention to new functions available in Freeze Mode, such as new trackball
functions.
3.
Press Freeze again to continue live image data acquisition.
Trackball
The trackball is used as computer mouse. It is surrounded by the trackball buttons. The
function of each button depends on the activated and on the current image acquisition mode.
Getting started
4-4
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4