Symbol
Meaning
Reference
Pictogram on Probe Care Card:
Do not immerse the probe into any liquid beyond the level
specified for that probe. Refer to the Instructions for Use of
the ultrasound system.
GE crafted symbol
Pictogram on Probe Care Card:
Describes precautions necessary to prevent the risk of
disease transmission or infections.
ISO 15223-1
Pictogram on Probe Care Card:
Describes precautions necessary to prevent the risk of injury
through electric hazards. Warning, electricity
IEC 60601-1; ISO
7010-W012
NRTL Classification Label
TUV crafted symbol
CE Conformity mark according to Medical Device Directive
93/42/EEC
93/42/EEC
Consult accompanying documents. This symbol advises the
user to consult the accompanying documents.
IEC 60601-1; ISO
7010-M002
Green dot on power cable plug Indicates that the power cable is hospital grade. Grounding
reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked “Hospital only”
or “Hospital grade”. Applicable depending on local regulatory
requirements.
ANSI/UL 498 and CSA
CAN/C22.2 No. 42-99
This indicates the maximum rated power consumption of the
system.
GE crafted symbol
Caution, consult accompanying documents. This symbol
advises the reader to consult the accompanying documents
for important safety-related information such as warnings and
pre-cautions that cannot be presented on the device itself.
IEC 60601-1; ISO
7000-0434A; ISO
15223-1:
No protection against ingress of water (system)
IEC 60529
Protected against water immersion - Immersion for 30
minutes at a depth of 1 meter. (ultrasound probes)
IEC 60529
Indicates a USB connector.
GE crafted symbol
Indicates a network connector. Computer network.
IEC 60417
Earth (ground) connection
IEC
60601-1:IEC60417-500
7
Product was refurbished / remanufactured by GE
ULTRASOUND KOREA, LTD.
GE crafted symbol