13.1 Safety conformance
The Voluson™ SWIFT / Voluson SWIFT+ conforms to the following standards and
regulations :
Conformance details
Emission:
CISPR11, EN 55011
Group 1 Class A
IEC/EN61000-3-2
Harmonic current emissions
IEC/EN61000-3-3
Voltage Change, Fluctuations and Flicker
Immunity:
IEC/EN61000-4-2
Electrostatic Discharge
IEC/EN61000-4-3
Radiated RF EM Fields
IEC/EN61000-4-4
Electrical Fast Transients and bursts
IEC/EN61000-4-5
Surges
IEC/EN61000-4-6
Conducted RF Fields
IEC/EN61000-4-8
Rated Power-frequency Magnetic Field
IEC/EN61000-4-11
Voltage Dips and Interruptions
Electrical safety:
IEC/EN60601-1
Mechanical safety:
IEC/EN60601-1
Thermal safety:
IEC/EN60601-1
Duty cycle:
100% on
Safety classification:
Class I & INTERNALLY POWERED ME EQUIPMENT, applied parts type BF acc. to
IEC/EN60601 incl. national deviations
Safety classification with ECG Class I, applied parts type CF acc. to IEC/EN60601 incl. national deviations
Ambient temperature:
•
18ºC to 30ºC resp. 64ºF to 86ºF (operation temp. of instrument)
•
-10ºC to 50ºC resp. 14ºF to 122ºF (storage and transport temp.)
Barometric pressure:
•
700 to 1060 hPa (operation condition)
•
700 to 1060 hPa (storage and transport condition)
Humidity:
•
30 to. 80% RH no condensation (operation condition)
•
10 to 90% RH no condensation ( Voluson™ SWIFT / Voluson SWIFT+ storage and
transport condition)
Humidity protection:
IPX0 no humidity protection
Maximum operating altitude:
3000m; depending on the properties of the connected electronic devices the maximum
operating altitude is limited to the altitude stated in the corresponding user manual of
the connected electronic device
Pollution degree:
2
Overvoltage category:
II
Material group:
IIIb
Light conditions
Natural & artificial light source (Bright light could impact readability of screen)
Type and model
Type: Voluson™
Model: Voluson™ SWIFT / Voluson SWIFT+
Technical Data/ Information
13-2
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4