Note
Biopsy Line buttons are greyed, if needle path was not calibrated once (Biopsy Setup).
Biopsy kit name and the “Biopsy Line” buttons are dependent on the selected probe.
Preparation for biopsy line adjustment
Read all safety relevant information before performing and programming a biopsy.
1. Water *
2. Biopsy-needle
* Water bath about 47˚ C
1.
Mount the desired biopsy guide to a probe and attach the needle.
2.
Connect the probe and select it.
3.
Press 2D on the user interface to activate B-Mode.
4.
Place the probe into a water bath (about 47˚C, set OTI to “Normal”) and display the exact
position of the needle on the active B image.
5.
Press the Menu on the left upper touch monitor.
6.
Press
System Setup
.
7.
Press
Biopsy Setup
.
8.
The Biopsy Setup Menu is displayed.
Note
A water bath alignment verification is also necessary before performing freehand biopsy
procedures.
5.4.3.1 Biopsy line adjustment for single angle biopsy guide
Storing a biopsy line
Figure 5-12 Biopsy Setup Menu: Single angle
biopsy guide
1.
Perform steps 1 to 8, see 'Biopsy setup'
2.
Select a biopsy kit.
3.
The biopsy line is shown on the monitor screen.
Probes and Biopsies
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
5-23
Содержание Voluson Swift
Страница 343: ......