Display modes:
B, XTD
Screen formats:
•
2D imaging: Single (B), Dual (B+B), Quad (B+B+B+B)
•
XTD View: Single (XTD)
13.10.2 M-Mode
Working modes:
M (conventional M-Mode) / AMM (Anatomical M-Mode)
Frequency range
1-18 MHz ( depending on the probe, 3 steps high, mid, low)
M Acoustic Power:
1 - 100
GAIN range:
+15 to -25 dB
M sweep speeds:
•
900 / 450 / 300 / 225 / 150 / 100 pixels/sec.;
•
26.44 / 13.22 / 8.81 / 6.61 / 4.40 / 2.94 cm/s in relation to system monitor
Review (memory times):
> 60 s (32MB)
Signal processing M:
•
Dynamic range: 1 to 12
•
Reject: 0 to 255
•
Enhance: 0 to 5
•
Gray maps: 18
•
Tint maps: 10
Display Modes:
•
M: 2D+M, 2D+M/CFM, 2D+M/HD-Flow™, 2D+M/PD
•
AMM: 2D+AMM, 2D/CFM+AMM/CFM, 2D/HD-Flow™+AMM/HD-Flow™
Screen formats:(window
arrangement)
•
2D+M and 2D+AMM: up/down (horizontal): three different sub formats 40/60,
50/50, 60/40%; left/right (vertical): 50/50%
•
2D+AMM+AMM: left//rt-up/rt-down: 50//25/25%
13.10.3 M-CFM-Mode
Acoustic MCFM Power
1-100
MCFM Color Maps
8 maps
Frequency range
1-18 MHz ( depending on the probe, 3 steps high, mid, low)
CFM Gain
+/-15 dB range, 0.1 dB steps
CFM Velocity Scale Range
PRF:100 Hz to 11 kHz
Wall Motion Filter
8 - 3000 Hz
Ensemble (color shots per
line)
8-16, step size 1
Gentle color filter
Smooth filter
Rise: 12 steps
Fall: 12 steps
CFM Spectrum Inversion
CFM Baseline Shift
17 steps
Pre-settable and independently adjustable B-, M and MCFM Gain
CFM Threshold
1 - 255 steps
Balance
25 - 255, step size 5
Artifact suppression
On/Off