2.15 Software Update
Software update for the unit may become available for download and installation through the
GE Service platform. When a software update is available a message icon is displayed on the
status bar.
Note
Software update through the GE service platform may not be available in all markets.
Info
Please contact a service technician for further assistance.
Software download and installation
1.
Press the standby button on the user interface.
The Shut Down dialogue window with the software
Download
button is displayed.
2.
Press
Download
.
The download process is started. The progression of the download process is displayed.
Download may be paused and can be resumed later.
3.
When the software download is completed, the
Download
button changes to
Install
.
Software installation may take about one hour.
4.
Press
Install
.
The system reboots automatically and the installation process is started. Do not interrupt
the installation. The system may reboot automatically several times during installation or
you may be prompted to restart the system.
When the installation is complete, the New Software Verification window is displayed.
5.
Perform a check for all features listed. Move the cursor over the feature name to get
information on how to check each feature.
If all features are
OK
the signature field is enabled.
6.
Enter your signature and press
Send
.
The system is ready.
If one feature gets “Failed”, the user will be prompted to reload the original software.
2.16 System messages
High system temperature
In case the temperature of the system is higher than usual, the following messages will be
displayed on the monitor screen. You can select the language in which the message is
displayed.
Caution
•
High system temperature reached! Further use of the system is possible. If this message
persistently appears, please contact your service agent.
•
High system temperature reached! If the system temperature stays high, the system will
shut down within 10 minutes.
Note
If the system needs to shut down, an additional message will be displayed in the message
area of the monitor screen:
High system temperature - system shutdown within [ ] minutes!
Safety
2-30
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
Содержание Voluson Swift
Страница 343: ......