Archive
•
Network
11.2.3.2.1 DICOM
DICOM is the abbreviation of Digital Imaging and Communications in Medicine. This is the
industrial standard for communication of images and other information between medical
devices on the network. Using the DICOM option, you can send or print images after
connecting your ultrasound equipment and PACS.
This dialog section is used to set up details of all of your DICOM target nodes (image servers).
Once you have set up a DICOM node properly, data can simply be transferred by selecting the
appropriate target node.
DICOM Configuration
Figure 11-16 DICOM Configuration
Controls
AE Title
Enter the AE (Application Entity) Title under which your DICOM application is known
to other DICOM applications (required). For information about the correct
AE Title
setting, please contact your DICOM network administrator.
Station Name
Enter the name of the hospital or institute.
Retry Count
Number of retries to establish a successful DICOM connection.
Retry Count Seq.
Retry count for sequential mode (only valid if
Send sequ. checkbox is checked for
the selected destination). If the end number of
Retry Count Seq. is reached and
sending was not successful then the “problem” data set is marked as “failed” in the
DICOM Transfer Queue Status and the system continues sending the next image
data
Retry Interval
Retry interval minutes.
Timeout (s)
Define a timeout interval.
Charater Set
Select desired language from a dropdown list consisting of ISO_IR 100 (Latin 1),
ISO-IR 101 (Eastern Europe), ISO_IR 126 (Greek), ISO_IR 138 (Hebrew), ISO_IR
144 (Cyrillic), ISO_IR 192 (UTF-8), ISO_2022 IR 87 (Japanese), GB18030
(Chinese).
Default
Default values of
Retry Count, Retry Count Seq., Retry Interval, Timeout (s) are set.
System Setup
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use