No
Type
Model name
Max. Length (m)
Type of Cable
Remark
7
Sector phased
array probe
3Sc-RS
Manufactured by GE Healthcare
1.95
Shielded
8
RIC5-9A-RS
Manufactured by GE Healthcare
2.4
Shielded
9
Volume convex
array probe
RAB2-6-RS
Manufactured by GE Healthcare
2.1
Shielded
10
RAB6-RS
Manufactured by GE Healthcare
2.1
Shielded
Table 2-4 List of all cables and maximum length of cables for Voluson™ SWIFT / Voluson SWIFT+
General precautions
Observe the following precautions:
•
The user is responsible for the safety of all persons in the vicinity of the ultrasound
system including the patient(s).
•
Get acquainted with the probes and the ultrasound system.
•
Misinterpretation of an ultrasound image can lead to false diagnosis.
•
The system is sensitive to shock and must be treated with care also if not in use.
•
Do not twist, kink or pinch cables. Excessive bending or stress on cables may result in
damage to its insulating properties and functionality.
•
Do not drop or subject the probe to any type of mechanical shock. Impact may
compromise probe operation, safety features or result in sharp edges that could damage
the protective sheath and / or injure sensitive tissue. Any damage caused by improper
use will void the warranty.
•
Installation and initial start-up of the system must be performed by authorized service
personnel.
•
For safety reasons, avoid handling fluids in the vicinity of the system.
•
Do not install software on the system that has not been released by GE, as this may lead
to erroneous function or function transfer and thereby decrease system performance.
•
If continuous operation is required even during power interruption, it is recommended
that the system is powered from an uninterruptable power source (UPS).
•
Each person who connects additional equipment to the signal input portion or signal
output portion configures a medical system, and is therefore responsible that the system
complies with the requirements of the system standard IEC 60601. If in doubt, consult
the technical service department or your local representative.
2.3 Electric installation
Local safety regulations may require an additional connection between the potential
equilibrium bolt and the building's grounding system.
Before connecting to power, ensure that the local mains voltage and frequencies correspond
to the indicated equipment configuration. See the voltage indicator on the power supply.
The minimum required house installation must have 10A.
Warning
Never use an adapter which does not ensure proper protective earth connection.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
Never remove or disconnect the protective earth.
Safety
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
2-11