Caution
After each use, inspect the lens, cable, and housing of the probe. Look for any damage that
would allow liquid to enter the probe.
Danger
If the probe is damaged, DO NOT place it into any liquid (e.g. for disinfection) and DO NOT
use it until it has been inspected and repaired/replaced by a GE Service Representative.
Probe Intermediate-Level Disinfection - Spray
1.
Put on a new pair of gloves and spray enough disinfectant solution to saturate a new
disposable low-lint wipe or cloth.
2.
Holding the probe near the strain relief, apply the dampened cloth to the patient
contacting lens. Wipe the probe from the lens to the strain relief, slightly rotating the
probe after each wiping pass.
3.
After the probe has been completely wiped, dampen a second wipe with disinfectant and
starting at the probe lens begin wiping the probe in a rotating motion moving down
towards the strain relief. Spray disinfectant directly on the recessed areas and ridges to
saturate.
Figure 5-7 Disinfecting the Probe Moving from Lens to Strain Relief
4.
Once the probe has been completely wiped, dampen a third wipe with disinfectant and
continue wiping the probe as needed to ensure the surface remains wet for the required
exposure time. Use as many wipes as needed and respray disinfectant on recessed
areas and ridges, to ensure all surfaces remain wet for the minimum required contact
time listed in the disinfectant manufacturer’s instructions for use.
5.
Thoroughly dry all surfaces of the probe using a soft, low-lint wipe or cloth, changing
wipes/cloths when necessary to ensure the probe is completely dry. Pat dry lens.
Visually inspect the probe to ensure all surfaces are dry. Repeat drying steps if any
moisture is visible.
6.
If the probe is not immediately reused, store the probe in a manner that will protect and
keep the probe from being recontaminated. This may be accomplished by placing the
probe in a storage cabinet with filtered air flow and/or by using a disposable storage
cover placed over the probe.
Probe Intermediate-Level Disinfection - Disinfectant Wipe
1.
Replace gloves. Holding the probe near the strain relief, apply the wipe to the patient
contacting lens. Wipe the probe from the lens to the strain relief, slightly rotating the
probe after each wiping pass.
Probes and Biopsies
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
5-11