Guidance and manufacturer´s declaration - electromagnetic immunity
The Voluson™ SWIFT / Voluson SWIFT+ is intended for use in electromagnetic environment specified below. The
customer or the user of the Voluson™ SWIFT / Voluson SWIFT+ should assure that it is used in such an environment.
Immunity Type
Equipment Capability
Regulatory Acceptable Level
Electromagnetic
environment - guidance
IEC/EN 61000-4-11
Voltage dips and
interruptions
< 5%UT (> 95% dip) for 0.5
cycle;
40%UT (60% dip) for 5
cycles;
70%UT (30% dip) for
25cycles;
< 5%UT (>95% dip) for 5
sec.
0 % UT; 0.5cycle, Phase:
0,45,90,135,180,225,270,31
5°
0 % UT; 1cycle, Phase: 0°
70 % UT; 25/30cycle,
Phase: 0°
0 % UT; 250/300 cycle
* Applicable deviations
standard IEC/EN
60601-2-37 (202.6.2.7)
Edition 3
< 5%UT (> 95% dip) for 0.5
cycle;
40%UT (60% dip) for 5
cycles;
70%UT (30% dip) for
25cycles;
< 5%UT (>95% dip) for 5
sec.
Edition 4
0 % UT; 0.5cycle, Phase:
0,45,90,135,180,225,270,31
5°
0 % UT; 1cycle, Phase: 0°
70 % UT; 25/30cycle,
Phase: 0°
0 % UT; 250/300 cycle
Mains power quality should
be that of a typical
commercial and/or hospital
environment. If the user of
the Model Voluson™
SWIFT / Voluson SWIFT+
requires continued operation
during power mains
interruptions, it is
recommended that the
Model Voluson™ SWIFT /
Voluson SWIFT+ be
powered from an UPS or
battery option.
IEC/EN 61000-4-8
Power frequency magnetic
field
3 A/m, 50/60Hz
30 A/m, 50/60Hz
Edition 3
3 A/m, 50/60Hz
Edition 4
30 A/m, 50/60Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial and/or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level
Safety
2-22
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4