Harm. Frequ.
Multi Frequency mode lets you downshift to the probe's
next lower frequency or shift up to a higher frequency.
Cine #
Select the desired Cine #.
6.2.5 Anatomical M-Mode (AMM)
Anatomical M-Mode gives you the ability to manipulate the cursor at different angles and
positions. The M-Mode display changes according to the position of the cursor.
Note
When AMM pre-mode is entered, the last used display format is activated. If the 2D user
program is changed, the display format saved in the use program is activated.
Using Anatomical M-Mode
1.
Select M-Mode.
2.
Touch the
AMM
button.
3.
Adjust the AMM-cursor in the 2D single image.
4.
Press
2D/M run
(left or right trackball button) to start AMM.
5.
Press the top trackball button to toggle between position and rotation of the AMM curser
(
pos
/
rot.
).
6.
Adjust or rotate the cursor accordingly.
7.
Press the small trackball buttons to select the different cursor lines
AMM1
or
AMM2
.
8.
Press
Freeze
.
AMM Main and Sub Menu
General M-Mode buttons are described in 'M-Mode'
.
AMM rot
AMM rotation
Note
The AMM angle is determined by the virtual MM cursor line and the AMM cursor line. At 0°
both lines have the same position. The maximum rotation is 180°. The max. CW/CCW rotation
angle is limited (max. horizontal position on the screen), so the AMM-cursor is mirrored when it
is not top/bottom oriented. During the angle rotation the image is updated in real time.
6.2.6 B-Flow
B-Flow helps to visualize complex hemodynamics and highlights moving structures or blood. It
is visually intuitive when viewing blood flow, for acute thrombosis, parenchymal flow and jets. It
is a realistic (intuitive) representation of flow information, allowing to view both high and low
velocity flow at the same time.
2D Mode
6-12
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4