9.5.1 Saving onto the Clipboard
Press the predefined Px-button (default: P1) on the user interface to save data onto the
clipboard.
Note
Pressing the Px-button without having started an exam will show a dialog:
You have to start an exam first!
•
Ok: Evokes the Current Patient Dialog to enter a patient. After entering the patient data
and returning to scan mode using the Start Exam button the image or cine will be stored
automatically.
•
Cancel: Cancels the dialog and returns to scan mode. No patient has been entered and
pressing the Px-button will lead to the same dialog.
The clipboard is filled from left to right and from top to bottom.
To scroll through the images on the clipboard:
1.
Swipe Up/Down with fingers.
2.
Drag Up/Down with the left trackball key and a trackball.
Saving reloaded data
Edited, reloaded images (Measurements, Annotations...) have to be saved to the clipboard
again, or else all changes will be lost when an other image is reloaded.
Depending on the system setup, saving of reloaded data follows certain rules:
•
Overwrite reloaded Image
: replaces the current reloaded image
•
Copy to the end of the clipboard
: saves an additional copy at the of the clipboard
•
Copy after the reloaded Image
: saves an additional copy after the reloaded image
Note
Existing dataset information will not be lost. (independent of the setting “Copy after the
reloaded Image”)
•
When a Volume Cine is saved as Singe Volume, the Volume Cine will not be overwritten.
The saved Single Volume will be saved to the end of the Clipboard.
•
If a 2D cine is saved as (single) image, the 2D Cine will not be overwritten. The image
will be saved to the end of the clipboard.
•
All screenshots are saved to the end of the clipboard.
•
Save TUI one by one does not replace the reloaded volume but will be saved to the end
of the clipboard.
9.5.2 Manipulating Files on the Clipboard
Use the trackball buttons or touch the buttons on the top bars to manipulate data on the
clipboard.
Reload from the clipboard
Double press the desired image with a touch gesture or the left trackball key.
Note
Disabled in Measure- and Annotation Text mode.
Note
Only one data set can be selected.
Repro function in clipboard
Press the Reproduce button to load and use the stored image presets.
Note
Disabled in Measure- and Annotation Text mode.
Archive
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
9-21
Содержание Voluson Swift
Страница 343: ......