GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
145
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
E18
GEN
None
IDC/SENO
Synchro. fail
A problem occurs in
synchronization
between Image
Detection Control
system and generator
during acquisition. May
be the image will be
fully acquired but in
that case, dose
parameters will be
missed (set to 0 in the
image).
Next exposure is
inhinited
1. Switch OFF the Gantry
2. Switch ON
3. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
4. Restart the system
5. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E19
GEN
None
IDC/SENO down,
DPS is OFF
Image Detection
Control system cannot
communicate with
Generator
The Dectector Power
Supply is then
powered OFF by the
system.
1. Switch OFF the Gantry
2. Switch ON
3. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
4. Restart the system
5. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E20
GEN
None
Check Main
Distribution Rack
Supply of AWS Cart
and Detector
environement circuits
is not authorized.
1. Switch OFF the Gantry
2. Switch ON
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E21
GEN
None
Detector supply
not supplied
Supply of Detector
circuits is not
authorized.
1. Switch OFF the Gantry
2. Switch ON
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E22
GEN
None
Conditioner not
supplied
Supply of detector
Conditioner circuits is
not authorized.
1. Switch OFF the Gantry
2. Switch ON
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E23
GEN
None
AWS cart not
supplied
Supply of WS Cart
circuits is not
authorized.
1. Switch OFF the Gantry
2. Switch ON
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.