Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
53
6-2
Automatic View Name Selection
6-2-1
The automatic view selection process
LMO
LMO
MLO or
SIO
MLO or
SIO
ML or LM
ML or LM
Laterality
Magnification
–45
_
Tube angles are measured from the vertical 0
_
position, with positive angles counter-clockwise
when facing the Gantry. The sketch shows the
tube at –4
5
_
.
FB
CC
Angulation
The operator selects the breast
laterality, Left or Right.
The system senses the presence or
absence of the magnification platform
The system selects a view name according to the
tube angle and the laterality. Default values for the
angles used are given in the table below.
The tube is at –4
5
_
and the selected laterality is R, so the inferior side
of the breast is against the receptor.
The system therefore selects
the view name SIO, and adds the laterality prefix R. Because the
magnification platform is present, it also adds the prefix M.
Automatically selected view name: R M SIO
Example:
Operator selected laterality: R.
Magnification platform: Present.
Tube angle: –4
5
_
.
180
_
0
_
+90
_
–90
_
6-2-2
Angles used in automatic view selection; default values
View Name
Selected
Tube Head to Left
Tube Head to Right
Selected
Minimum
Maximum
Minimum
Maximum
CC
–10
_
+10
_
Only one setting for CC
MLO or SIO
+11
_
+79
_
–79
_
–11
_
ML or LM
+80
_
+100
_
–100
_
–80
_
LMO
+101
_
+169
_
–169
_
–101
_
FB
+170
_
+180
_
–180
_
–170
_
Note that these values can be changed by the user: