CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
161
Message on AWS screen
Corrective action
Explanation
AA
10
”Shutdown Process started
...”
The AWS shutdown process has
been started because a UPS low
battery condition was detected
(battery power must be available
to guarantee safe operation and
shutdown in the event of power
failure).
1. Stop system usage and wait for the end
of the shutdown sequence.
2. Check the mains power supply.
3. Check (on UPS panel) that the batteries
are charging. Do not attempt to restart the
system until the batteries have recharged.
4. If the problem persists, contact GEMS
Field Service.
AA
11
System is down.
System will power off within 30
seconds: do NOT reboot now!
(if system does not turn off,
switch off UPS manually)
The message is displayed at the
end of the normal shutdown
sequence, when the UPS is
about to shut down, to prevent
users from trying to reboot.
No action is normally required.
If the system fails to shut down completely
(the AWS screen continues to display the
message after one or two minutes), switch
it off by pressing the
UPS
OFF
button at
the front of the Generator Cabinet.
AA
12
”System not ready, exam
forbidden”
A system component is not yet
ready (after system stsrt-up).
Wait until system initialization is complete
before requesting an acquisition.
AA
13
UPS battery needs replacing.
Please call service
The UPS selftest has detected
that the battery requires
replacement.
Stop system usage and contact GEMS
Field Service.
AA
14
UPS output overload.
Please call service
An UPS overload condition has
been detected.
Stop system usage and contact GEMS
Field Service.
BB
These messages relate to boot/reset events:
BB
1
”AWS – IDC communication
failure, exam forbidden.”
Communication between the
AWS and the acquisition system
was not established at boot time,
or has been lost during
application.
1. Click
OK
on the pop-up window
2. Restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.
BB
2
AWS/SENO communication
failure, exam forbidden
Communication between the
AWS and the Gantry system was
not established at boot time, or
has been lost during application.
1. Click
OK
on the pop-up window
2. Reset the gantry (power OFF/ON),
and/or restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.
BB
3
AWS not functional, exam
forbidden
One of the system components
did not boot (or did not reset)
correctly.
1. Click
OK
on the pop-up window
2. Restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.