CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
149
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
I14
GEN
None
AWS was not
ready
AWS system was not
ready to perform an
image acquisition.
1. Press the prep button again.
2. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
5. Restart the system
6. If the problem persists, note ALL
messages and inform GEMS Field
Service.
I15
GEN
None
Console locked
The control console is
locked due to service
access (detector
calibration or IQ tool is
in progress). Actions
on right–hand section
of control controle are
still available.
Wait end of service access.
If no service in progress power
OFF/ON the DMR.
I16
GEN
None
Collimator locked The diaphragm control
is locked because
detector calibration is
in progress.
Wait end of calibration.
If no calibration in progress power
OFF/ON the DMR.
I17
GEN
None
AWS/SENO
comm. lost
AWS system cannot
communicate with
Generator.
May be a service PC
laptop is connected in
place of AWS or may
be the AWS cart is
powered OFF for
service need.
1. Wait end of service operation if
there is one else wait 1mn to allow
the system to recorver itself,
2. If the problem persists, power OFF
the whole system and wait for the
end of the shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
I18
GEN
None
Collimator/Consol
e locked
The control console
and the diaphragm
control are locked due
to service access
(detector calibrationis
in progress). Actions
on right–hand section
of control controle are
still available.
Wait end of service access.
If no service in progress power
OFF/ON the DMR.