CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
154
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
S01
GEN
None
Aborted exposure
Exposure in progress
has been aborted
because the Exposure
Button was released
by operator before the
end of exposure.
Switch OFF Gantry, then switch ON.
Start exam again.
S02
GEN
None
Exposure aborted
by HV arcing
Exposure was aborted
because more than
eight x-ray tube arcing
events occurred.
Switch OFF Gantry, then switch ON.
Start exam again.
S05
GEN
None
Wrong selection
of kV/track couple
Selected kV/focal track
is not enabled (kV
values are too low in
the Rh focal track).
Change the configuration.
S07
GEN
None
Exposure aborted
by IDC
in OM
1. Switch OFF the Gantry
2. Switch ON
3. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
4. Restart the system
5. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S08
GEN
None
Required power
is too high
Power requested in
Manual Mode is too
high.
Reduce kV or mAs values.
S09
GEN
None
No compression
detected
Breast is not
compressed.
Breast compression is mandatory.
S10
GEN
None
Generator cooling
mn
Generator requires
xxxx
minutes to cool.
Wait for cooling period (or lower mAs
values in Manual Mode).
S11
GEN
None
Tube cooling mn
Tube requires
xxxx
x
minutes to cool.
Wait for cooling to terminate (or
lower mAs values in Manual Mode).
S12
GEN
None
Configuration not
allowed
Incompatible settings
of magnification, grid,
and focal spot have
been selected.
Select compatible settings (refer to
the Image Acquisition chapter).
S13
GEN
None
Backup mAs < 50
exposure
inhibited
Maximum mAs < 50.
Exposure inhibited.
Wait for cooling to terminate (or
lower kV values in AEC Mode).