CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
157
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
S82
AWS
None
S82 Image
detection failure
One of the image
detection coponent is
failed.
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S83
AWS
None
S83 Detector
environment not
OK
The environment of
the detector is not
correct to allow
acquisition.
1. Wait for completion of system
setting.
2. If this X–ray inhibition stays
displayed more that 10 mn
switch OFF the whole system and
wait for the end of the shutdown
sequence.
3. Restart the system.
4. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S84
AWS
None
S84 Image
quality is
degraded
The temperature drifts
slowly from the optimal
value specified for best
Image Quality.
Check image quality on current
exam.
If the problem persists, note ALL
messages and inform GEMS Field
Service.
S85
AWS
None
S85 Detector
environment
failure
The environment of
the detector is failed.
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S86
AWS
None
S86 AWS/IDC
Communication
lost
AWS system cannot
communicate with
Image Detection
control system but the
communication should
be recorved itself.
1. Wait that the Image Detector
control system finished to recorver
itself.
S87
AWS
None
S87 AWS/IDC
Communication
failure
AWS to Image
Detection control
system communication
is failed.
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.