CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
146
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
E24
GEN
None
Failure before
exposure
Current image
acquisition aborted
before exposure due
to failure in the
equipment
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E25
GEN
None
AWS/SENO
Synchro. fail
A problem occurs in
synchronization
between AWS system
and generator during
acquisition.
Next exposure is
inhinited
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence
2. Restart the system
4. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E26
GEN
None
AWS/SENO
comm. failure
AWS system cannot
communicate with
Generator.
1. Switch OFF the Gantry
2. Switch ON
3. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
4. Restart the system
5. If the problem persists, note ALL
messages and inform GEMS Field
Service.
E35
GEN
None
SYSTEM
SHUTDOWN
STARTED
A power OFF of the
whole system has
been started.
The generator will
power off the supply of
Detector, Gantry and
Generator in fiew
minutes.
Stop system usage and wait the end
of the shutdown sequence.
E36
GEN
None
CART PWR
FAILURE – SYST
SHUTDOWN
The system shutdown
has been processed
because an UPS error
has been detected or
communication with
UPS haas been lost.
1. Stop system usage and wait the
end of the shutdown sequence.
2. Check the main power supply.
3. Check (on UPS panel) that the
batteries are charging. Do not
attempt to restart the system until the
bateries have recharged.
4.If the problem persists, note ALL
messages and inform GEMS Field
Service.
E50
GAN
Failure
Software Failure
Gantry software error. Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and error codes
(ESPECIALLY THE 6 DIGITS OF
E50 MESSAGES) and contact
GEMS Field Service.