CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
144
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
E06
GEN
None
Software error
(gen CPU)
An error occurred in
the Generator CPU
software.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and inform GEMS Field
Service.
E08
GEN
None
Power supply
failure
Supply of generator
power circuits is not
authorized.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E10
GEN
None
Heating failure
track 1
Molybdenum focal
track is unserviceable.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can still work with rhodium focal
track.
E11
GEN
None
Heating failure
track 2
Rhodium focal track is
unserviceable.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can still work with molybdenum
focal track.
E12
GEN
None
Checksum error
(gen)
Calibration parameters
have been modified.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E13
GEN
None
Gen./Arm
communication
fail
Generator cannot
communicate with
arm.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact
GEMS Field Service.
E14
GEN
None
Generator
CPU/INTG failure
A problem has been
reported by the
Generator CPU.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E15
GEN
None
Console
communication
fail
Generator cannot
communicate with
Control Console.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.