GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
151
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
I28
GEN
None
Max wait for
same exposure
mn
It is necessary to wait
the displayed time to
be able to repeat the
same exposure.
Wait the required amount of time to
repeat exposure. If an exposure is
attempted before this time, an abort
may possibly occur (there is a safety
margin to ensure proper operation).
You can ignore the message if a
different, less dense, breast is to be
imaged next.
I29
GEN
None
Excessive mAs
for this mode
mAs
The amount of mAs
necessary to repeat
the same exposure is
too high in the mode
selected.
Switch to a different mode or to a
different filter.
I31
GEN
None
Abort: Image
signal too low
I32
GEN
None
Abort: Image
signal too high
I33
GEN
None
Abort: No AOP
zone found
I34
GEN
None
Thickness out of
AOP range
I46
GEN
None
System Setting in
progress
AWS set-up in
progress (loading of
Flat Field map) after a
new filter was
selected.
No action required.
This message is displayed for two
seconds after selection of a new
filter.
I50
GAN
Lock
Jammed
Lock Jammed
LOCK PROBLEM:
In spite of command,
lock release does not
operate.
Try again to release lock.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
I51
GAN
Hot Lock
Lock Overheated
Lock temperature too
high (used for too long
period and/or too
frequently)
Wait for lock to cool.
I52
GAN
Arm
Unlocked
Tube Housing
Arm Unlocked
Tube housing arm in
unstable position
because lock is badly
latched.
Pivot arm to latch correctly.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
I54
GAN
Wait...
Power Limitation
Heat protection for 27
V supply. Elevator,
rotation, and light
centering device lock
are blocked.
Leave equipment idle for a while.
If the problem continues, note ALL
messages and contact GEMS Field
Service.