CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
155
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
S14
GEN
None
Maximum mAs
exceeded
Maximum mAs set at
installation are
exceeded. (This
message appears only
if a value below the
maximum possible
mAs value was set at
installation).
Lower mAs values.
S20
GEN
None
Acq. Abort, image
can be lost
Exposure has been
fully performed but
image may be lost
because an error
occurred in the image
chain detection or
processing.
1. Press RESET.
2. check image quality if image is
displayed
2. Reposition the patient.
3. Start exam again
2. If the problem persists, power OFF
the whole system and wait for the
end of the shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
S21
GEN
None
Waiting for AWS
boot
Communication with
AWS system is not yet
established.
1. Wait AWS boot and login is
finished.
2. If message persists 10mn after the
AWS login power OFF the whole
system and wait for the end of the
shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
S22
GEN
None
Waiting for IDC
boot
Communication with
Image Detection
Control sysetm is not
yet established.
1. Wait Image detection Control boot
is finished.
2. If message persists 10mn power
OFF the whole system and wait for
the end of the shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
S23
GEN
None
Laterality is not
selected
You have not selected
the laterality of the
breast to perform
acquisition. Exposure
is inhibited
Select laterality on the Control
Console. If you are imaging a test
object, the laterality chosen is not
important.
S25
GEN
None
Tube housing
over temperature
The temperature of the
tube housing is too
high (> 65
°
C).
Wait for the tube housing
temperature to fall.