CHAPTER 14 – ERROR MESSAGES
1.
ERROR MESSAGES
Any operational anomalies found by the system during Senographe 2000 D operation are signalled by
Error Messages
. These may be displayed on the Column Readout and on the Control Console or in a
pop-up box on the AWS screen. They may be simple information messages requiring no action, or they
may indicate error or fault conditions requiring action from the operator and/or GEMS Service
engineers.
Note that messages displayed on the Gantry begin with three identification characters, which are
followed by the message itself. On the Control Console, the message is followed by seven further
identification characters. Thus, for example, the Column Readout might display
I76 FAILURE
while the
Control Console displays
I76 SENSOR FAILURE 158/011
. If it necessary to call GEMS Field Service
about an error, you should quote the entire message, complete with the identification characters.
The tables below list error messages which may occur, with explanations and recommended actions.
Table 1 lists error messages displayed on the Gantry: control console and/or column.
Source of the message is mentionned:
GAN = issued from Gantry, GEN = from generator, AWS = from Acquisition workstation.
Table 2 lists error messages displayed on the AWS screen.
Table 1 – Error Messages displayed on the Gantry
Msg.
No.
Source
Message
on Column
Readout
Message on
Control Console
Explanation
Corrective Action
E 01
GEN
None
Generator failure
Problem reported in
HV section of
generator.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E02
GEN
None
Focus bias failure
Problem reported in
tube focus bias
circuits.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E03
GEN
None
Rotor failure
Problem reported in
anode starter.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E04
GEN
None
Arm fatal failure
Generator detects
error in dialog with
arm.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
E05
GEN
None
Failure during
exposure
Current exposure
aborted due to failure
in equipment.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and inform GEMS Field
Service.