Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
150
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
I19
GEN
None
IDC/SENO
comm. lost
Image Detection
Control system cannot
communicate with
Generator.
If the communication is
not recorver in few
minutes, the sysetm
will be in failure mode
(error E19)
1. Wait that the Image Detector
control system finished to recorver
itself.
I21
GEN
None
Heating failure
track 1
Molybdenum focal
track is out of use.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can continue using rhodium
focal track.
I22
GEN
None
Heating failure
track 2
Rhodium focal track is
out of use.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can continue using molybdenum
focal track.
I23
GEN
None
Console
communication
fail
Generator-Control
Console connection is
faulty.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I24
GEN
None
Line power
interruption
Micro power cut
occurred on line
supply.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
I25
GEN
None
Recumbent
patient view
names
Operator has selectec
a view name for
recumbent patient.
If you don’t want to run acquisition
with this selected view name, enter
again to the set–up/Proj. menu
without valid to come back to normal
view name selection.
Else the view name selection will
come–back to normal selection after
the next exposure.
I27
GEN
None
Generator CPU
battery error
CPU backup battery is
discharged.
Contact GEMS Field Service.
DO NOT REMOVE POWER FROM
EQUIPMENT.