CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
148
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
E86
GAN
Chksum
Error
Arm checksum
error
Content of saved
memory has been
modified. Calibration
may be false.
Compression is
inhibited.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
I05
GEN
None
Exposure aborted
by mAs level
Exposure aborted
because not enough
mAs available to
continue exposure.
1. Press RESET.
2. Reposition the patient.
3. Start exam again
4. If the problem continues, note ALL
messages and contact GEMS Field
Service.
5. Nevertheless, you can continue to
use the system with manual
acquisition.
I06
GEN
None
Abort: Detect sig.
out range
Exposure aborted
because brightness
received by detector
are not in range to
continue exposure.
1. Press RESET.
2. Reposition the patient.
3. Start acquisition again
4. If the problem continues, note ALL
messages and contact GEMS Field
Service.
5. Nevertheless, you can continue to
use the system with manual
acquisition.
I07
GEN
None
Selected filter not
allowed : Al
Filter selected by the
operator not allowed.
The Senograph
automatically selects
the next correct filter.
1. Press ”Filter selection” to have the
correct filter selected.
I12
GEN
None
Maximum mAs =
Maximum available
mAs are equal to
xxx
.
Information message only.
(If mAs exceeded, cell aborts
exposure at 2 or 3 mAs if entire
exposure cannot be made.)
I13
GEN
None
IDC was not
ready
Image Detection
Control system was
not ready to perform
an image acquisition.
1. Press the prep button again.
2. Switch OFF the Gantry
3. Switch ON
4. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
5. Restart the system
6. If the problem persists, note ALL
messages and inform GEMS Field
Service.