Table 2 – Error messages displayed on the AWS monitor
Message on AWS screen
Explanation
Corrective action
AA
These messages relate to power on/off events:
AA
1
Could not connect to the
remote provider
The specified remote host could
not be found. It may be
temporarily off-line, its address
may have changed or been
incorrectly entered, the
application on the remote
provider may not be running, or
there may be a network error.
1. Try again after a short wait.
2. If the message is repeated, check that
the server address is correct and that the
remote application is running.
3. If the problem persists, contact GEMS
Field Service.
AA
2
Do you really want to power
off the system: yes or no?
This message is displayed as
part of the normal shutdown
sequence.
No corrective action required (click
Yes
or
No
as required)
AA
3
”End of remote connection in
progress”
A GE Insite connection is being
closed.
No action is required; you can now use the
system.
AA
4
”Image detection failure,
exam forbidden”
An image detection subsystem is
not ready for acquisition after
start-up.
1. Wait to see if the problem clears.
2. If the problem persists, contact GEMS
Field Service.
AA
5
”Initialization not complete,
please wait”
A system initialization process is
in progress.
Wait until system initialization is complete.
AA
6
”No Ups Power Backup”
A UPS error has been detected
or communication with the UPS
has been lost.
1. Stop system usage and shut down the
system.
2. Check the mains power supply.
3. Check (on UPS panel) that the batteries
are charging. Do not attempt to restart the
system unless the batteries have been
recharged.
3. Restart the system.
3. If the problem persists, contact GEMS
Field Service.
AA
7
”Senograph failure, exam
forbidden”
A Senographe (DMR) subsystem
is not ready for acquisition after
start-up.
1. Wait to see if the problem clears.
2. If the problem persists, contact GEMS
Field Service.
AA
8
”Shutdown Process started
...”
The AWS shutdown process has
been started because a UPS
primary power failure longer than
10 seconds has been detected.
1. Wait for the end of the shutdown
sequence.
2. Check mains power supply.
3. Restart the system when power OK.
4. If the problem persists, contact GEMS
Field Service.
AA
9
”Shutdown Process started
...”
The AWS shutdown process has
been started because
communication with the UPS
was lost for more than 30
seconds.
1. Stop system usage and wait for the end
of the shutdown sequence.
2. Restart the system
3. If the problem persists, contact GEMS
Field Service.