Electrical testing
FUJIFILM SonoSite products tested
The following equipment has been tested:
l
Vevo MD Imaging System
l
Vevo MD Imaging System transducers: UHF22, UHF48, UHF70
l
Foot switch
Electrical safety testing
The system complies with the following medical equipment standards related to
electrical safety as follows:
l
CAN/CSA-C22.2 No. 60601-1:14 Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance
l
EN 60601-1:2006/A1:2013 Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance
l
ANSI/AAMI ES 60601-1:2005/(R)2012/A1:2012, C1:2009/(R)2012 and
A2:2010/(R)2012 Medical electrical equipment-Part 1: General requirements for
basic safety and essential performance
l
JIS T 0601-1:2012 Medical electrical equipment-Part 1: General requirements
for basic safety and essential performance
Electromagnetic compatibility
The system complies with the electromagnetic compatibility (EMC) limits as per the
following standards:
Electrical safety
355
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...