16 Obtain
a new cloth moistened with a compatible disinfectant or disinfectant wipe.
Compatible disinfectant materials are listed in the table
17 Disinfect
the transducer body by wiping the surface.
18 Ensure
the transducer stays wet with the disinfectant for the wet contact time
listed in the table
"Compatible cleaners and disinfectants"
This may
require reapplication to achieve the required wet time
.
♦ For example, if you are using Sani-Cloth Bleach it was a wet contact time of 4
minutes. Wipe the transducer cable and body until thoroughly wet. If the transducer
starts to dry before 4 minutes is reached, use a wipe and re-wet the transducer.
Continue the process of re-wetting with a new wipe until a total wet time of 4 minutes
is reached.
19 Allow
the transducer to air dry in a clean, well ventilated space.
20 Examine
the transducer and cable for damage, such as cracks or splitting where
fluid can enter.
If damage
is evident, do not use the transducer, and contact
FUJIFILM SonoSite or your local representative.
DONE
330
Cleaning and disinfecting
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...