control
b. Tap any location along the depth scale and the depth control will automatically
move to the new position—once the depth is moved, the value will update
2
As you adjust the depth, the depth value also appears in the lower left corner of the
clinical display.
Note:
The depth value in the lower left corner of the clinical display is always the
total acquired depth of the non zoomed image. When you zoom, this value stays the
same.
Gain and TGC controls
Gain refers to amplifying the intensity of the returning sound waves on the screen
display. Increasing the gain brightens the image. Decreasing the gain darkens the
image. Use the gain control to adjust the visual intensity of the signal.
As you adjust the master gain control the Time Gain Compensation (TGC) controls
are displayed. During image acquisition in any 2D-based imaging mode, each slider
adjusts the ultrasound signal to compensate for minor attenuation as it returns through
deeper situated tissue.
Each slider adjusts the return signal across a specific depth band. The top slider
adjusts the return signal across the area closest to the transducer face. The bottom
slider adjusts the return signal across the area furthest from the transducer face.
Slide the control to the right to boost the signal and brighten the image data in that
horizontal band, and left to attenuate the signal and darken that band.
The current value is displayed above the control as you adjust it.
A
TGC
button is also available on the imaging controls bar on the right of the control
panel. Tap the
TGC
button to enable the TGC sliders. The TGC sliders will remain on
the control panel until they are changed or if you tap anywhere else on the control
panel.
220
Scanning
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...