Continue to
"Part 2 of 2: High-level disinfecting of transducer for semi-critical
Part 2 of 2: High-level disinfecting of transducer for semi-critical contact
WARNING:
Wear the appropriate personal protective equipment (PPE)
recommended by the chemical manufacturer, such as eyewear and gloves.
High-level disinfectants can cause harm to the patient if not completely
removed from the transducer. Follow the manufacturer’s rinse instructions
to remove chemical residue.
Caution:
Do not skip any steps or abbreviate the disinfecting process in any way.
Do not use unapproved disinfecting agents.
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended disinfectants. Using a non-
recommended disinfecting solution or incorrect solution strength can damage or
discolor the transducer and void the transducer warranty.
Transducer
Disinfectant
Temperature
Disinfectant
soak duration
Clean water
soak time
UHF22
UHF48
UHF70
Cidex
25°C (77°F)
45 minutes
3 x 1 minute
Compatible high-level disinfectants for semi-critical contact
1 Ensure
you have cleaned the transducer as seen in
system and transducer, disinfecting of system for semi-critical contact"
Step A: Semi-critical contact cleaning and disinfecting of system and
transducer
323
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...