Acoustic output measurement
Since the initial use of diagnostic ultrasound, the possible human biological effects
(bio effects) from ultrasound exposure have been studied by various scientific and
medical institutions. In October 1987, AIUM ratified a report from its Bio effects
Committee (Bio effects Considerations for the Safety of Diagnostic Ultrasound, J
Ultrasound Med., Sept. 1988: Vol. 7, No. 9 Supplement). The report, sometimes
referred to as the Stowe Report, reviewed available data on possible effects of
ultrasound exposure. Another report, “Bio effects and Safety of Diagnostic
Ultrasound,” dated January 28, 1993, provides more current information.
The acoustic output for this ultrasound system has been measured and calculated in
accordance with “Acoustic Output Measurement Standard for Diagnostic Ultrasound
Equipment” (NEMA UD2-2004), and “Standard for Real-Time Display of Thermal and
Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment” (NEMA
UD3-2004).
In Situ, derated, and water value intensities
All intensity parameters are measured in water. Since water does not absorb acoustic
energy, these water measurements represent a worst case value. Biological tissue
does absorb acoustic energy. The true value of the intensity at any point depends on
the amount, type of tissue, and the frequency of the ultrasound passing through the
tissue. The intensity value in the tissue,
In Situ
, has been estimated by using the
following formula:
In Situ
= Water e
[-(0.023alf )]
where:
In Situ
=
In Situ
intensity value
Water = Water intensity value
e = 2.7183
a = attenuation factor (dB/cm MHz)
384
Appendices
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...