START
Semi-critical cleaning and disinfection
1 Put on
gloves, for protection from cleaning and disinfecting materials. Change
gloves as needed throughout the procedure.
2 Turn off
the system by pressing the
Power button
.
3 Turn off
the main power switch located on the bottom of the system near the back,
where the power cord connects.
4 Unplug
the power cord from the outlet.
5 Unplug
the network cable, if applicable.
6 Disconnect
and remove the transducer from the system.
7 Remove
the transducer sheath (probe cover), if applicable.
8 Obtain
a compatible cleaner. Compatible cleaning materials are listed in the table
"Compatible cleaners and disinfectants"
Caution:
Make sure the solution does not get into any of the connector ports.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces if not using a wipe.
9 Clean
the transducer cable and transducer body to remove any visible gel, debris,
or bodily fluids.
♦ Starting from the transducer connector end, clean the transducer cable and
transducer body in a downward motion toward the scanhead. Wipe from clean areas
to the soiled areas to avoid cross-contamination.
Caution:
Do not wipe the transducer connector.
Step A: Semi-critical contact cleaning and disinfecting of system and
transducer
321
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...