WARNING:
To avoid the risk of electrical shock: Use only properly
grounded equipment. Shock hazards exist if the power supply is not
properly grounded.
Grounding reliability can only be achieved when equipment is connected to
a receptacle marked “Hospital Only” or “Hospital Grade” or the equivalent.
This equipment must be connected only to a supply mains with protective
earth.
Do not let any part of the system (including the power supply, or power
supply connector), touch the patient. Do not touch the power supply and the
patient at the same time.
Do not touch the signal input/output connectors on the back of the
ultrasound system or the system transducer connector when the transducer
is disconnected.
Do not connect the system’s power supply to a Multiple Socket-Outlet
(MSO) or extension cord.
Before using the transducer, inspect the transducer face, housing, and
cable. Do not use the transducer if the transducer or cable is damaged. Turn
the system off when cleaning. Do not use any transducer that has been
immersed beyond the specified cleaning or disinfection level.
See
"Compatible accessories and peripherals"
on page 361 recommended
by FUJIFILM SonoSite, including the power supply. Connection of
accessories and peripherals not recommended by FUJIFILM SonoSite could
result in shock. Contact FUJIFILM SonoSite for a list of accessories and
peripherals available from or recommended by FUJIFILM SonoSite.
WARNING:
Inspect the AC power cords, cables, and plugs on a regular
basis. Ensure that they are not damaged to avoid the risk of electrical shock
and fire hazard.
352
Appendices
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...