Operation
Automatic setting
Alarm limits can be automatically adapted to current measured values or set values.
1. Open the
Alarms
>
Limits
dialog.
2. To adjust the alarm limits for
one
individual parameter, touch and confirm one of
the
Autoset
buttons (5).
To adjust the alarm limits for
all
parameters, touch and confirm the
Autoset all
button (4).
The function for adjusting all alarm limits can be called up directly with the
Autoset
limits
button in the main menu bar.
Only use the automatic adjustment when measured values and set values are
stable.
The lower alarm limit for the
xMAC
level is also adjusted during automatic setting,
see page 118.
Configuration and algorithm, see page 169.
7.9.2 Activating
and deactivating CO
2
alarms
Prerequisite: The device has the "Integrated patient-gas measurement module"
option.
The CO
2
monitoring (alarms for
inCO
2
,
etCO
2
, and
CO
2
apnea) can be activated or
deactivated.
1. Open the
Alarms
dialog.
2. Touch the
Settings
tab (1).
3. For
CO
2
alarms
, touch the button (2):
On
: Alarms are activated.
Off
: Alarms are deactivated.
Or
►
Use the
CO
2
alarms off
button in the main menu bar to activate or deactivate
the alarms.