Tests
For passive anesthetic gas scavenging:
1.
WARNING
Risk of overpressure
If the overpressure valve in the passive AGS or the scavenging hose is blocked,
overpressure in the breathing circuit and in the patient’s lungs will occur.
►
Take care that the scavenging hose does not become blocked.
►
Perform a visual inspection of the overpressure valve for damage and soiling.
Check that the scavenging hose is run correctly. The hose must not be blocked.
2. Perform the visual inspection of the overpressure valve (4).
8.4.6 Accessories
8.4.6.1 Soda
lime
►
Make sure that the soda lime does not need to be exchanged. Change the soda
lime if it is discolored or when its maximum period of use has been reached, see
page 75.
With Infinity ID function:
– Absorbers of type Infinity ID CLIC Absorber will be detected automatically.
The replacement date will be set automatically.