Troubleshooting
!! 255
Breathing system failure
The breathing system is not
correctly assembled or
installed.
Check patient condition.
Check the breathing system.
Verify that the sensor port for
the O
2
sensor is sealed.
Check if the piston dia-
phragm is correctly inserted.
If the problem persists,
replace the breathing sys-
tem.
The breathing system is
faulty.
Replace the breathing sys-
tem.
If ventilation is impaired, use
a manual resuscitator.
!!! 255
Breathing system temp.
high?
The breathing system
warmer is faulty.
Check the inspiratory breath-
ing gas temperature as close
to the Y-piece as possible.
Remove breathing system
cover.
Use longer inspiratory hose.
Open the system setup and
turn off the breathing system
warmer.
If necessary, proceed as fol-
lows: -Turn off the device
with the main switch, then
ventilate the patient with the
breathing bag. Or -Discon-
nect the patient, then use a
manual resuscitator.
!
100
Breathing system warmer
failure
The breathing system
warmer is faulty.
Check breathing circuit for
condensation. Increase
fresh-gas flow if necessary.
Call Dräger.
!! 100
Cardiac bypass mode still
active?
A significant minute volume
was measured during car-
diac bypass mode.
Deactivate the cardiac
bypass mode.
Use "ALARM RESET" to
downgrade alarm priority.
!
100
Central Air supply high
The central supply pressure
is high. The gas delivery
might fail.
Check central supply.
!
255
Central Air supply low
Central supply pressure is
low.
Check central supply.
!
100
Central N
2
O supply high
The central supply pressure
is high. The gas delivery
might fail.
Check central supply.
!
255
Central N
2
O supply low
Central supply pressure is
low.
Check central supply.
Priority
Alarm
Cause
Remedy