Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
163
Alarms
10.6
Activation of alarms after breath detection
Product variants with inspiratory O
2
measurement or integrated patient-gas
measurement module
In the
Man / Spon
mode, the alarms for
Minute volume low
and
Apnea (no flow)
are only activated after spontaneous breaths have been detected.
Product variants with integrated patient-gas measurement module
If still no breaths have been detected after leaving the
Standby
or
Pause
modes,
the breathing gas is monitored with regard to an O
2
concentration that is too low or
an anesthetic gas concentration that is too high. At the same time, the message
Waiting for respiratory phases
is displayed in the CO
2
waveform.
Once 2 breaths have been detected, the message disappears and only then are the
O
2
, CO
2
, N
2
O, and anesthetic gas alarms active.
Cardiac bypass mode
still active?
If a minute volume of
>50 % of the suggested
value is measured after
CBM mode has been
activated for
>60 seconds
---
---
Fresh gas low or leak-
age
---
Breathing bag empty
With an additional
"Apnea (no pressure)"
alarm and at least one of
the following alarms:
"Apnea (no flow)", "Tidal
volume not achieved" or
"Airway pressure not
achieved"
Alarm
Priority
Priority
Priority
Note
(Low)
Caution
(Medium)
Warning
(High)