Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
277
Technical data
T
0...90
<45 s (RR
≥
6 /min)
<105 s (RR <6 /min)
Leakage minute volume
MVleak
Range
0 to 40 L/min
Accuracy
±25 % of measured value or ±0.2 mL/min
(the larger value applies)
Resolution of displayed value
0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV
≥
1 L/min)
Low-flow wizard
Standardization
The following data are standardized to STPD
conditions.
Range
0 to 8 L/min
Accuracy
±25 % of measured value or ±100 mL/min
(the larger value applies)
Respiratory rate
Total RR
Spontaneous
RRspon
Mandatory
RRmand
Range
0 to 100 /min
Accuracy
±10 % or ±1 /min (the larger value applies)
Resolution of displayed value
1 /min
T
0...90
<45 s (RR
≥
6 /min)
<105 s (RR <6 /min)
Dynamic compliance
Dynamic compliance
Cdyn
Mean dynamic compliance
Cdyn mean
Range
0 to 200 mL/hPa (mL/cmH
2
O)
Accuracy for gas measurement with PGM
±15 % or ±1 mL/hPa (mL/cmH
2
O) (the larger
value applies)
Accuracy without PGM
±40 % or ±3 mL/hPa (mL/cmH
2
O) (the larger
value applies)
Information
The values may be distorted by spontaneous
breathing, thus reducing the measurement
accuracy.
Resolution of displayed value
0.1 mL/hPa (mL/cmH
2
O)
Resistance
R
Range
0 to 100 hPa/L/s (cmH
2
O/L/s)
Accuracy for gas measurement with PGM
±30 % of the measured value or ±3 hPa/L/s
(cmH
2
O/L/s) (the larger value applies)
16.13 Measuring
systems
and displays (continued)