Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
47
Assembly and preparation
4.1.1 Electrical
safety
Ambient conditions
If the device is operated or connected to the mains power supply at ambient
temperatures above 35 °C (95 °F), the battery cannot be charged properly. The
power supply out of the battery may be limited. As a result, the patient could be put
at risk.
►
Do not expose the device to temperatures above 35 °C (95 °F) on a permanent
basis.
Mains power supply
If the device is connected to a power socket with incorrect mains voltage or a power
socket without a protective ground conductor, an electric shock may occur.
Connecting the device to auxiliary power sockets can lead to an increased leakage
current. This may result in an electric shock and the user and patient may be put at
risk.
►
Connect the device only to power sockets with correct mains voltage and a
protective ground conductor.
►
Do not connect the device to additional power socket strips.
Battery supply
If the battery is not sufficiently charged, it may not be possible to maintain operation
for long enough if the mains power supply fails. If there is a mains power supply
failure, devices connected to the auxiliary power sockets are not supplied from the
internal battery. As a result, the patient could be put at risk.
►
Before first operation or after storage, charge the battery for at least 8 hours.
►
Check the functional integrity of the battery by performing regular inspections.
Auxiliary power sockets
The connection of devices to auxiliary power sockets can lead to an increased
leakage current. If the protective ground conductor of one of these devices fails, the
leakage current may rise above the permissible values. This may result in an
electric shock or device failure and the patient and user may be put at risk.
►
Do not connect additional power socket strips to the integrated auxiliary power
sockets on the device, but rather to separate power sockets on the wall.
►
Have the leakage current checked by service personnel.
►
If the permissible value is exceeded, use separate power sockets on the wall
instead of the auxiliary power socket on the device.
►
Do not connect high-frequency surgery equipment to the auxiliary power sockets
of the anesthesia machine.
►
When making a connection, follow the manufacturer's instructions for all
connected devices.