![Dräger Atlan A300 Скачать руководство пользователя страница 57](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728057.webp)
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
57
Assembly and preparation
5. Put the breathing system cover (5) in place and click it into position.
4.3.3
Establishing the mains power supply
The mains voltage must correspond to the voltage range indicated on the rating
plate on the rear of the device.
To protect from inadvertent disconnection of the power cable, the power inlet of the
device is secured with a guard plate.
1.
WARNING
Risk due to incorrect mains voltage or missing protective ground
conductor
If the device is connected to a power socket with incorrect mains voltage or a
power socket without a protective ground conductor, an electric shock may
occur.
►
Connect the device only to power sockets with correct mains voltage and a
protective ground conductor.
WARNING
Risk of electric shock
If the device is connected to an additional power socket strip, this may lead to
increased leakage current. The leakage current may exceed the permissible
values.
►
Do not connect the device to additional power socket strips.
►
Do not connect additional power socket strips to the integrated auxiliary
power sockets on the device, but rather to separate power sockets on the
wall.
Plug the power plug into the power socket.
The power plug must be readily accessible so that the power supply to the
device can be interrupted quickly in the event of a device malfunction.
2. Using the symbols on the status display, check that the power supply is
established.
3. Turn on the device.
4.3.4 Charging
the
battery
The internal battery will automatically start charging as soon as the device is
connected to the mains power supply.
36505