Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
63
Assembly and preparation
4.5.2.3 Handling
O
2
gas cylinders
►
WARNING
Risk of explosion
When pressurized, O
2
is self-igniting in combination with oil or grease.
►
Do not oil or grease the gas cylinder valve or the pressure reducer on the O
2
cylinder. Do not touch with oily or greasy fingers.
The gas cylinder valves must only be opened and closed slowly. Do not use any
tools with the screw connection variants.
Have service personnel replace any leaky or stiff gas cylinder valves.
4.5.3 Fitting
the
vaporizers
The device can be operated with vaporizers which have a Dräger Auto Exclusion
plug-in adapter or a Selectatec plug-in adapter. Dräger recommends using only
vaporizers that are listed in the list of accessories and have been tested.
The vaporizers used must conform to the ISO 8835-4 or ISO 80601-2-13 standard.
The illustration shows vaporizers of the type Dräger-Vapor 3000.
Connecting the vaporizers
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause leakage. This can cause the fresh-gas
delivery to be too low or contaminate the ambient air. Patient and user can be
endangered.
►
Make sure that the connected vaporizers are hanging vertically.
►
When using D-Vapor vaporizers, make sure that the power cable is not pinched.
►
After mounting the vaporizers, perform a leakage test.
1. Set all the vaporizers upright and securely on the plug-in adapter.